APP下载

多参数监护仪检定结果的不确定度评定

2024-05-30陈淑荣

品牌与标准化 2024年2期
关键词:不确定度评定血氧血压

【摘要】多参数监护仪能够对病人的多种生命体征参数进行实时监测和记录,在临床上主要用于监测危重症患者的生命体征参数。本文依据多参数监护仪检定规程,对心电、无创血压、血氧三个模块的测量结果进行了不确定度的评定,可为相关从业者提供参考。

【关键词】多参数监护仪;不确定度评定;血压;血氧

【DOI编码】10.3969/j.issn.1674-4977.2024.02.072

Uncertainty Evaluation of Verification Results of Multi-Parameter Monitors

CHEN Shurong

(Beijing Xicheng Institute of Metrology, Beijing 100055, China)

Abstract: The multifunction patient monitoring instrument can monitor and record various vital sign parameters of patients in real time. It is mainly used to monitor vital sign parameters of critically ill patients in clinic. According to the verification regulation of multiparameter monitor, the uncertainty of the measurement results of ECG, non-invasive blood pressure and blood oxygen was evaluated. It can provide reference for relevant practitioners

Keywords: multi-parameter monitor; uncertainty evaluation;precautions; blood pressure; blood oxygen

1概述

1.1测量依据

JJG 1163—2019《多参数监护仪》和JJF 1059.1—2012《测量不确定度评定与表示》。

1.2环境条件

温度:25.6℃;湿度:65.8%。

1.3检定标准器具

生命体征模拟仪,型号:NIM-PMS-1,厂家:中国计量科学研究院,编号:M2106009。

1.4测量对象

医用多参数监护仪,型号:ePM 10M,厂家:深圳迈瑞生物医疗电子股份有限公司,编号:AD1-19024178。

2电压测量误差的不确定度评定

2.1测量模型

2.2標准不确定度分析

2.2.1由测量结果引入的标准不确定度

测量不确定度主要来源于电压幅度的读数分散性,采用钢直尺测量电压幅度值,测量重复性和直尺分辨力都会造成读数的分散性,应取二者中数值大的。

1)由重复性引入的相对标准不确定度

在重复性条件下进行10次测量,采用A类方法进行评定,记录的幅值长度为:10.1 mm、10.1 mm、10.0 mm、10.1 mm、10.1 mm、10.0 mm、10.1 mm、10.1 mm、10.0 mm、10.0 mm。

10次测量结果的平均值为xˉ= 10.06 mm。

【参考文献】

[1]多参数监护仪:JJG 1163—2019[S].

[2]测量不确定度评定与表示:JJF 1059.1—2012[S].

[3]郑冬梅.多参数监护仪检定装置不确定度评定[J].机电信息,2023(3):73-76.

【作者简介】

陈淑荣,女,1988年出生,工程师,一级注册计量师,硕士,研究方向为医学计量器具的计量检定。

(编辑:李加鹏)

猜你喜欢

不确定度评定血氧血压
稳住血压过好冬
智能血氧饱和度监测系统设计与实现
血压偏低也要警惕中风
血压的形成与降压
基于STM32血氧心率检测仪的研制
预应力混凝土用钢绞线拉伸强度试验结果不确定度的评定
基于血氧模拟前端AFE4490的无创血氧测量模块设计
乳腺血氧与多普勒联合检查在乳腺癌早期筛查中的应用
控制血压刻不容缓