Monte Carlo Simulation of in situ Gamma-Spectra Recorded by NaI (Tl) Detector in the Marine Environment
2015-03-31WANGYimingZHANGYingyingWUNingWUBingweiLIUYanCAOXuanandWANGQian
WANG Yiming, ZHANG Yingying, WU Ning, WU Bingwei, LIU Yan, CAO Xuan, and WANG Qian
Monte Carlo Simulation ofGamma-Spectra Recorded by NaI (Tl) Detector in the Marine Environment
WANG Yiming1), ZHANG Yingying2), *, WU Ning2), WU Bingwei2), LIU Yan2), CAO Xuan2), and WANG Qian2)
1),,266061,..2),,266001,..
To develop a NaI (Tl) detector for in situ radioactivity monitoring in the marine environment and enhance the confidence of the probability of the gamma-spectrum analysis, Monte Carlo simulations using the Monte Carlo N-Particle ( MNCP ) code were performed to provide the response spectra of some interested radionuclides and the background spectra originating from the natural radionuclides in seawater recorded by a NaI (Tl) detector. A newly developed 75mm×75mm NaI (Tl) detector was calibrated using four reference radioactive sources137Cs,60Co,40K and54Mn in the laboratory before the field measurements in seawater. A simulation model was established for the detector immersed in seawater. The simulated spectra were all broadened with Gaussian pulses to reflect the statistical fluctuations and electrical noise in the real measurement. The simulated spectra show that the single-energy photons into the detector are mostly scattering low-energy photons and the high background in the low energy region mainly originates from the Compton effect of the high energy γ-rays of natural radionuclides in seawater. The simulated background spectrum was compared with the experimental one recorded in field measurement and they seem to be in good agreement. The simulation method and spectra can be used for the accurate analysis of the filed measurement results of low concentration radioactivity in seawater.
Monte Carlo simulation; marine radioactivity monitoring; NaI (Tl) detector; response spectra; background spectra
1 Introduction
The radioactive monitoring in the marine environment has been using the method of in situ sampling, subsequent processing and analyzing in the laboratory in China. This complicated handling procedure always takes two or three days to get the quantitative results. It is impossible to timely and effectively monitor the radioactivity in the marine environment, not to mention providing the pollution warning. Thus there is a need to develop an autonomous underwater detector for the continuous and long-term radioactive monitoring in the marine environment.
NaI (Tl) scintillation crystal detectors are the most common forradioactivity measurements in seawater due to their good efficiency, high reliability, low power consumption and low cost. There are a few NaI (Tl) detectors and even marine radioactive monitoring networks abroad (Tsabaris,2008;Wedekind,1999;Osvath, 2005;Caffrey, 2012). In China, NaI (Tl) detector has been widely used for the rapid measurements of natural and artificial radionuclides on land. The research on NaI (Tl) detector for in situ radioactive monitoring in the marine environment just gets started (Pan., 2014; Su., 2010 ).
The γ-ray spectra, recorded by NaI (Tl) detector, are complicated because of the interactions between γ-photos and substances in seawater. Especially, there is a high background in the low energy region mainly originating from the Compton effect of natural radionuclides such as40K with high concentrations in seawater. These make it difficult to discriminate the few artificial radionuclides such as137Cs and further to quantitatively calculate them. Thus, to improve the spectrum analysis, it is important to study the radioactive spectra in the marine environment, including the background and certain artificial radionuclides. But it is very difficult to carry out the test experiment in the laboratory, since there are a rather limited number of single-energy, gamma-emitting radionuclides soluble in seawater, and there is no seawater tank big enough to imitate the open sea environment and put some artificial radionuclides into it. The big volume of the contaminated seawater is very difficult to handle.
For the development and application of in situ NaI (Tl) detector for the marine environment in China, Monte Carlo simulations were performed in the present work using the MNCP code to study the γ-ray spectra of in situ radioactive measurement. The energy and efficiency calibrations of the detector developed were described. The background and response spectra were simulated for the radionuclide discrimination and calculation. In order to test the reliability of these simulations, experimental spectra of the detector were deduced from the field measurements and used to compare with the simulation results.
2 Simulation Model and Method
Forradioactive monitoring in the marine environment, an underwater NaI (Tl) detector was developed. The detector, consisting of a 75mm×75mm NaI (Tl) crystal connected with a photomultiplier and integrated electronic circuits with a density of 3.667×10−3gmm−3, was packaged in a watertight cylindrical enclosure made from stainless steel. A simulation model was established for the NaI (Tl) detector. The photomultiplier as well as the integrated electronic circuits and cables of the detector were discarded, and so the detector was simplified to be merely a scintillator crystal and placed in the centre of the seawater tank in the model. The model geometries are shown in Fig.1. The attenuation of γ-ray due to stainless steel enclosure around the crystal is considered in the model for better simulation. To simulate the measurement of the detector immersed in seawater, the seawater with standard composition and density of 1.025gcm−3is even- ly distributed in the tank modeled with radius of=100, 200,…,1500mm.
Fig.1 Measurement simulation model of NaI (Tl) detector immersed in seawater.
To reflect the statistical fluctuations and electrical noise in the measurement, the γ-ray spectra recorded by the NaI (Tl) detector were broadened with Gaussian pulses in the simulation. The broadening parameters were deduced from the experimental tests of the detector developed.
3 Calibration and Measurement
The NaI (Tl) detector had to be calibrated before its deployment in the seawater. In the laboratory, the detector was energy-calibrated and efficiency-calibrated using four reference radioactive sources137Cs,60Co,40K and54Mn. In the seawater, there is a certain radionuclidewith the different activities, and the counting rates of its photopeak are also different. A relative counting rate is defined as the photopeak counting rate of the above-mentioned radio- nuclide per unit activity. The relative counting rate can be used to represent the marine detection efficiency.
Some experimental tests were carried out in the laboratory and field measurements were completed near the Eight Gap Qingdao port. In the field measurements, the detector was deployed with a crane into seawater as shown in Fig.2, the water depth being 4 to 8m. The detector was positioned 3m below the water level in order to eliminate the measurement interferences from the seabed and cosmic radiation. It was connected to a notebook for realtime measurement and data collection.
Fig.2 Detector ready to immersed into seawater in field measurement.
4 Response Spectra Simulation
The full width at half maximum (FWHM) of137Cs,60Co and40K were all taken from the experimental tests and that of131I was given by the fitting curve of FWHM. All their photopeaks were taken from the experimental tests in the laboratory. Then the broadening parameters of Gaussian pulses for the radionuclides were deduced for their response spectrum simulations. The curves of the calculated response spectra of NaI (Tl) detector are shown in Fig.3. It can be seen that the single-energy photons into the detector are mostly scattering low-energy photons. The calculation mode used in the simulation is mode P and the particle number is 1e8.
Fig.3 The calculated response spectra of NaI (Tl) detector. a) for131I; b) for137Cs; c) for60Co; d) for40K.
5 Background Spectrum Simulation
Monte Carlo simulations using the MNCP code were performed to produce the background spectrum induced by the natural radionuclides in the marine environment. There are 37 radionuclides from the natural series of232Th,238U and235U in seawater. Among them, the γ-rays of several radionuclides such as214Bi,214Pb,212Bi,208Ti,228Ac,.have considerable influences on in situ radioactive monitoring of NaI (Tl) detector. There are also some long-lived radionuclides existing independently in seawater. Among them,40K is the most important. Then a number of strong γ-rays, weighted by their relative intensity, were selected for the background spectrum simulation. The background spectrum is the sum over all response spectra of strong γ-rays. By adjusting the relative contri- bution of these spectra and getting the best fit to the real spectrum (taken with NaI (Tl) of the detector developed near the Eight Gap Qingdao port), the background spectrum simulated, as shown in Fig.4, is in good agreement with the experimental one. There is a high background in the low energy region, which makes it difficult to discriminate the few artificial radionuclides mainly with low energies and low concentrations.
1.平行避险线(不应从漂浮物体引出,如灯浮,而应从固定物,如灯塔,灯桩等);2.海图的更换;(两海图接图—接图点、船位等);3.定位的方法和时间间隔(驾驶员最容易疏忽);4.明显的导航和雷达物标;5.禁止进入区(不鼓励过多地标绘“禁止进入区”);6.避险线和避险方位;7.叠标、导标和导航线;8.重要的潮流和海流;9.安全航速和必需的航速变化;10.最小富裕水深;11.应开启回声测深仪的船位;12.(距危险物的)安全距离;13. 锚位宽余量;;14.意外事件计划;15.放弃进港计划的最后位置;16. VTS和报告点等。
6 Conclusions
There are many differences between in situ radioactive monitoring in the marine environment and that on land. In order to improve the performance of underwater NaI (Tl) detector and produce more reliable results in the concentration of low level radioactivity in seawater, Monte Carlo simulations using the MNCP code were performed in the present work to generate the response spectra of some interested radionuclides and the background spectrum in seawater recorded by NaI (Tl) detector. It can be seen that the single-energy photons into the detector are mostly scattering low-energy photons, and the high background in the low energy region mainly originates from the Compton effect of the high energy γ-rays of natural radionuclides in seawater. The comparison of the simulated background spectrum with the experimental one shows promising results for the simulation code as well as for the spectrum analysis.
The development of in situ radioactive monitoring in the marine environment using NaI (Tl) detector just start- ed in China. We anticipate that sustained research into the Monte Carlo simulation ofmarine radioactivity measurement will be conducted and the results will be used for the detector development and applications. Based on the simulation results, the spectrum analysis algorithm is the next step.
Acknowledgements
We acknowledge the financial support from the International Science & Technology Cooperation Program of China (No. 2013DFR90220), National Natural Science Foundation of China (No. 41206076) and Qingdao Applied Basic Research Project (NO. 14-2-4-94-jch). The discussion with Dr. Christos Tsabaris in Hellenic Center for Marine Research is greatly appreciated. The authors are sincerely grateful for the comments and suggestions of the anonymous reviewers.
Bagatelas, C., Tsabaris, C., Kokkoris, M., Papadopoulos, C. T., and Vlastou, R., 2010. Determination of marine gamma activity and study of the minimum detectable activity (MDA) in 4pi geometry based on Monte Carlo simulation., 165: 159-168.
Caffrey, J. A., Higley, K. A., Farsoni, A. T.,and Smith, S., 2012. Development and deployment of an underway radioactive cesium monitor off the Japanese coast near Fukushima Dai-ichi., 111: 120-125.
Casanovas, R., Morant, J. J., and Salvadó, M., 2013. Implementation of gamma-ray spectrometry in two real-time water monitors using NaI (Tl) scintillation detectors., 80: 49-55.
Graaf, E. R., Limburg, J., Koomans, R. L., and Thijs, M., 2011. Monte Carlo based calibration of scintillation detectors for laboratory andgamma ray measurements., 102: 270-282.
Grujic, S., ÐorCevic, I., Milosevic, M., and Kozmidis-Luburić, U., 2013. Monte Carlo simulation of GM probe and NaI detector efficiency for surface activity measurements., 58: 45-51.
Jehouani, A., Ichaoui, R., and Boulkheir, M., 2000. Study of the NaI (Tl) efficiency by Monte Carlo method., 53: 887-891.
Nedjadi, Y., Spring, P., Bailat, C.,Decombaz, M., Triscone, G., Gostely, J. J., Laedermann, J. P., and Bochud, F. O., 2007.
Primary activity measurements with 4pi NaI (Tl) counting and Monte Carlo calculated efficiencies., 65: 534-538.
Osvath, I., Povinec, P. P., Livingston, H. D., Ryan, T. P., Mulsow, S., and Commanducci, J. F., 2005. Monitoring of radioactivity in NW Irish Sea water using a stationary underwater gamma-ray spectrometer with satellite data transmission., 263: 437-440.
Pan, X. B., Zhu, Y. N., Yu, X. R., Quan, L., and Qu, G. W., 2014. Monte Carlo Modeling of NaI(Tl) detectors for nuclides137Cs and131I in seawater., 5: 177-181 (in Chinese with English abstract).
Peeples, J. L., and Gardner, R. P., 2012. Monte Carlo simulation of the nonlinear full peak energy responses for gamma-ray scintillation detectors., 70: 1058-1062.
Su, G. H., Zeng, Z., Cheng, J. P.,and Zhu, L., 2010. Monte Carlo study on detection efficiency of seawater measuringgdetectors., 30: 451-459.
Tsabaris, C., 2008. Monitoring natural and artificial radioactivity enhancement in the Aegean Sea using floating measuring systems., 66: 1599-1603.
Vitorelli, J. C., Silva, A. X., Crispim, V. R., Fonseca, E. S., and Pereira, W. W., 2005. Monte Carlo simulation of response function for a NaI (Tl) detector for gamma rays from 241 Am/Be source., 62: 619-622.
Vlachos, D. S., and Tsabaris, C., 2003. Monte Carlo simulation of NaI (TI) gamma-spectra in sea water., 69: 370-372.
Vlastou, R., Ntziou, I. T., Kokkoris, M., Papadopoulos, C. T., and Tsabaris, C., 2006. Monte Carlo simulation of γ-ray spectra from natural radionuclides recorded by a NaI detector in the marine environment., 64: 116-123.
Wedekind, C., Schilling, G., Gruttmuller, M., and Becker, K., 1999. Gamma-radiation monitoring network at sea., 50: 733-741.
Yalcin, S., Gurler, O., Kaynak, G., and Gundogdu, O., 2007. Calculation of total counting efficiency of a NaI (Tl) detector by hybrid Monte-Carlo method for point and disk sources., 65: 1179-1186.
Yi, C. Y., and Hah, S. H., 2012. Monte Carlo calculation of response functions to gamma-ray point sources for a spherical NaI (Tl) detector., 70: 2133- 2136.
(Edited by Ji Dechun)
10.1007/s11802-015-2841-4
(December 25, 2014; revised February 10, 2015; accepted February 19, 2015)
. E-mail:triciayyz@163.com
ISSN 1672-5182, 2015 14 (3): 471-474
© Ocean University of China, Science Press and Springer-Verlag Berlin Heidelberg 2015
猜你喜欢
杂志排行
Journal of Ocean University of China的其它文章
- Effects of Waterborne Cu and Cd on Anti-oxidative Response,Lipid Peroxidation and Heavy Metals Accumulation in Abalone Haliotis discus hannai Ino
- A General Overview of the Typical18Frontal-Ventral-Transverse Cirri Oxytrichidae s. l. Genera (Ciliophora, Hypotrichia)
- Application of a Delay-Difference Model for the Stock Assessment of Southern Atlantic Albacore (Thunnus alalunga)
- Invasion and Morphological Variation of the Non- Indigenous Barnacle Chthamalus challengeri (Hoek, 1883) in Yangshan Port and its Surrounding Areas
- The Burial of Biogenic Silica, Organic Carbon and Organic Nitrogen in the Sediments of the East China Sea
- Hydrochemistry of the Natural Low pH Groundwater in the Coastal Aquifers near Beihai, China