Terahertz toroidal dipole metamaterial sensors for detection of aflatoxin B1
2024-03-25JianweiXu徐建伟ShoujianOuyang欧阳收剑ShouxinDuan段守鑫LinerZou邹林儿DanniYe叶丹妮SijiaYang杨思嘉andXiaohuaDeng邓晓华
Jianwei Xu(徐建伟), Shoujian Ouyang(欧阳收剑), Shouxin Duan(段守鑫),Liner Zou(邹林儿), Danni Ye(叶丹妮), Sijia Yang(杨思嘉), and Xiaohua Deng(邓晓华)
1Department of Physics,School of Physics and Materials Science,Nanchang University,Nanchang 330031,China 2Department of Measuring and Controlling Technology and Instrument,Institute of Advanced Manufacturing,Nanchang University,Nanchang 330031,China
3Institute of Space Science and Technology,Nanchang University,Nanchang 330031,China
Keywords: terahertz,metamaterial,toroidal dipole,aflatoxin B1
1.Introduction
Terahertz(THz)waves,ranging from 0.1 THz to 10 THz,have attracted much attention since they have significant scientific and technological potential in many fields.In particular, THz biosensing has become a promising tool in biological research due to its label-free, real-time andin situdetection advantages.[1-3]Metamaterials exhibit unique electromagnetic characteristics that would not be possible in natural and synthetic materials.[4]So far, the development of metamaterials lies in diverse aspects such as invisibility cloaks,[5]superlenses,[6]filters,[7]modulators[8]and sensors.[9]In particular,metamaterial sensors have attracted much attention due to their small size, ease of integration and excellent performance.Fan and colleagues have done a lot of excellent work in this field.[10-14]For instance,they realized thermal denaturation sensing, concentration sensing and identified types of aqueous solutions of proteins using a flexible twisted duallayer metamaterial with reflective THz time-domain polarization spectroscopy.They also used an all-dielectric metamaterial with functionalized gold nanoparticles to realize immunosensing of target biomolecules at nanolevel concentrations.Later, they successively used a Pancharatnam-Berry metamaterial sensor and electromagnetically induced transparency (EIT) metamaterial sensor to realize the identification of chiral enantiomers of amino acids.Recently, they have conducted further in-depth research and implemented the recognition of chiral enantiomers of amino acids in an aqueous environment using napole metamaterial sensors.Now,high-sensitivity, real-time monitoring sensing can be realized in metamaterial systems because a high sensitivity and high quality (Q)-factor can be obtained by designing exotic metamaterial structures.For example, metamaterial sensors based on Fano resonance[15-19]and EIT[20-23]were shown to have a sharp resonance that improved performance.[24]Most recently, a new type of metamaterial constructed from toroidal dipole structures was proposed in which coupling between the toroidal dipole and free space is relatively weak, leading to the effective localization of the electromagnetic energy and extreme sensitivity to environmental disturbance.[25-29]
In this paper, mirror symmetric THz metamaterial sensors consisting of periodic toroidal dipole structures are designed.Theoretical analyses and simulations show that the sensitivity andQ-factor can be as high as 337.5 GHz per refractive index unit(RIU)and 26.6,respectively.These values of sensitivity andQ-factor are superior to those of most current metamaterial systems.[30-34]The toroidal dipole sensor is then experimentally fabricated for detection of aflatoxin B1(AFB1).The results demonstrate that for an analyte of AFB1 solution the minimum measurable concentration can be as low as 0.001µg·ml-1.
2.Design and fabrication
A schematic view of the proposed periodic toroidal dipole metamaterial is shown in Fig.1(a).Here the substrate is polyimide (PI) film with thicknesshsub= 50 µm, and the permittivityεand loss tanδof the PI are set as 3.1 and 0.05,respectively.[35]The unit cell of the toroidal dipole is demonstrated in Fig.1(b) and consists of two semicircular copper rings with inner diameterR=18 µm and a copper rod with lengthL=54µm and widthw=6µm.The gap between the rod and ring,the thickness of the copper structure and the period of the unit cell areg=3µm,h=200 nm andP=70µm,respectively.When a THz wave in TE mode(electric field is oriented alongy)is vertically incident to the toroidal dipole,as shown in Fig.1(c),the bilaterally symmetric surface currentjcan be excited and the loop magnetic fieldM,which connects head to tail, is confined to a small area.That is, the toroidal dipole vectorTis formed in theydirection.We used CST Microwave Studio to simulate the designed structure; all parameters and simulation settings are described in Appendix A.
In our experiment the toroidal dipole metamaterial sample is fabricated as follows.First,the silicon wafer and PI film are successively cleaned by ultrasonic shock with acetone and ethanol.Then,the PI film is pasted onto the silicon wafer using polydimethylsiloxane with a ratio of main to curing agents of 10:1.Then,a 200 nm thick copper film is deposited on the PI film by an electron beam evaporation coating system.Finally, the toroidal dipole metamaterial sample is obtained, as shown in Fig.1(d),by photolithography and etching.
Fig.1.THz toroidal dipole metamaterial sensor.(a) Schematic diagram of the proposed sensor for AFB1 sensing.(b)Unit cell of toroidal dipole metamaterial with thickness of PI hsub =50 µm, inner diameter R=18 µm, length L=54 µm, width w=6 µm, gap g=3 µm,thickness of copper h=200 nm and period P=70 µm.(c) Diagram of toroidal dipole generation.(d) Microscopic image of metamaterial sample in the experiment.
3.Results and discussion
3.1.Transmission of toroidal dipole metamaterial
The transmission of the toroidal dipole metamaterial by simulation (red curve) and experimental measurement (black curve)is illustrated in Fig.2(a).The simulation was performed using CST Microwave Studio with normalized TE incidence,and the experimental measurements were made with a THz time-domain spectrometer (THz-TDS).The simulated transmission shows a sharp dip at 1.33 THz and is consistent with the experimental result at 1.3389 THz.To explore the properties of the toroidal dipole in more depth,surface current density andxozplane magnetic field at 1.33 THz are simulated by CST and demonstrated in Figs.2(b)and 2(c),respectively.This verifies that the bilaterally symmetric surface currentj(Fig.2(b)) and the loop magnetic fieldH(Fig.2(c)) are excited at the resonance dip, The equivalent charges formed at the metal ends of the bilateral symmetrical half rings are respectively recorded as (q1,-q1), (q3,-q3), and the corresponding electric dipole moments are respectively expressed asP1andP3.The equivalent charge at the end of the middle metal rod is recorded as (q2,-q2), and the equivalent electric dipole moment is expressed asP2, which is opposite to the direction ofP1andP3and suppresses and cancels them.The bilaterally symmetric surface currentjcan be equivalent to a pair of opposite parallel magnetic dipolesM, which also cancel each other, but they will form a toroidal dipole vectorT, as shown in Fig.1(c), when connected head to tail.The loop magnetic field in Fig.2(c)also shows that the metamaterial has the characteristics of a toroidal dipole externally at the resonance frequency.In addition,theQ-factor can be defined asQ=f0/FWHM to quantitatively describe the sensing performance,wheref0is the resonance frequency and FWHM is full width at half maximum.Here,Q=26.6 can be obtained at the 1.33 THz resonance dip.
Fig.2.(a) Simulated and experimental transmission spectra of the toroidal dipole metamaterial.(b)Surface current density j and(c)xoz plane magnetic field vortex H of the toroidal dipole metamaterial at 1.33 THz.
3.2.Biosensor based on the toroidal dipole metamaterial
Based on the properties of the toroidal dipole with sensitivity to geometrical parameters, THz metamaterial biosensors can be designed as shown in Fig.1(a).Specifically, for a biosensor system with the analyte on the Cu metamaterial,the transmissions of the metamaterial obtained by changing the thicknesstof the analyte at fixed refractive indexn=1.6 are shown in Fig.3(a).This shows that the resonance dip occurs at 1.33 THz fort= 0 µm.Fort= 3 µm, 6 µm,9 µm, 12 µm and 15 µm, the resonance dips have redshifts of 163 GHz,190 GHz,200 GHz,204 GHz and 205 GHz,respectively.In Fig.3(b), the dependence of those frequency shifts and the fitting ontis illustrated.It shows that whentexceeds 12 µm the resonance frequencies are approximately steady.The transmission of the metamaterial when changing the refractive indexnof the analyte from 1 to 1.8, with fixedt= 12 µm, is depicted in Fig.3(c).Compared with 1.33 THz whenn=1, the resonance dips also undergo redshifts of Δf= 67 GHz, 136 GHz, 204 GHz and 269 GHz when refractive indexn=1.2, 1.4, 1.6 and 1.8, respectively.In particular, the dependence of Δfonnand its fitting curve is illustrated in Fig.3(d).Here, Δfhas a good linear relationship withn.We note that the sensitivityScan be defined asS=Δf/Δn, where Δnis the variation of refractive indexn.Here,S=337.5 GHz/RIU can be obtained for the toroidal dipole structure,implying that it can be used as a highly sensitive refractive index sensor.
Fig.3.Biosensor based on the toroidal dipole metamaterial.(a)Transmission and(b)frequency shift of the sensor upon changing the thickness t of the analyte with fixed refractive index n=1.6.(c)Transmissions and(d)frequency shifts obtained by changing the refractive index n of the analyte from 1 to 1.8 with fixed t=12µm.
3.3.Detection of AFB1 by the toroidal dipole metamaterial
Aflatoxins (AFs) are the most dangerous group of mycotoxins due to their highly toxic, mutagenic, teratogenic and carcinogenic properties.[36]Among the aflatoxins, AFB1 is the most hepatocarcinogenic and hepatotoxic and occurs as a contaminant in a variety of foods.[37]So far, many methods, including traditional high-performance liquid chromatography,[38-40]enzyme-linked immunosorbent assay[41-43]and liquid chromatography coupled to tandem mass spectrometry,[44-46]have been proposed to detect AFB1.However, they are generally time-consuming, require expensive equipment and dangerous chemicals and are laborintensive.In comparison, detection based on THz metamaterial structures could be quick and convenient.In particular,when different concentrations of AFB1 are dropped on our toroidal dipole metamaterial AFB1 analyte layers of different thicknesses are formed.The higher the concentration of AFB1 solution,the thicker the AFB1 layer.Variations of THz transmission spectra can be observed for highly sensitive sensing.The preparation and detection process for different concentrations of AFB1 solutions are detailed in Appendix B.We further expanded the differences in the AFB1 analyte layer by adding and drying the AFB1 solution three times in the experiment, which can improve the sensitivity and detection limit while reducing the impact of experimental error.This method can also increase the number of samples added according to the actual situation to achieve more sensitive detection.
Figures 4(a) and 4(b) separately show the sample and microscopic image of the toroidal dipole metamaterial with 100 µg·ml-1of AFB1.The transmission and corresponding variations for different AFB1 concentrations are demonstrated in Figs.4(c) and 4(d), respectively.Here, for very low concentrations,the frequency of the resonance dip shows no distinct change, but its dip amplitude varies.Specifically,when the concentration varies from 0 to 0.001 µg·ml-1, and 0.01µg·ml-1,the dip amplitudes move from 0.1904 to 0.203,and 0.2093, respectively.For slightly larger concentrations,both resonance frequency and dip amplitude shift.In detail, for AFB1 concentrations of 0.1 µg·ml-1, 1 µg·ml-1,10 µg·ml-1and 100 µg·ml-1, amplitudes of 0.2179, 0.226,0.2384 and 0.2527, respectively, and redshifts of 10.1 GHz,20.1 GHz,27.7 GHz and 37.6 GHz,respectively,are obtained.By referring to the method of estimating the limit of detection(LOD)in Ref.[47],we determined that the LOD of AFB1 under our experimental conditions was less than 0.001µg·ml-1,which is lower than the requirement for AFB1 residue of most agricultural products and ensures food safety.
Fig.4.Detection of AFB1 by the toroidal dipole metamaterial: (a) sample and (b) microscopic images of toroidal dipole metamaterial with drops of 100µg·ml-1 of AFB1.(c)Transmission spectra of toroidal dipole sensors with different concentrations of AFB1.(d)Dependence of frequency shift and amplitude variations of the transmission dips on the concentration of AFB1.
4.Conclusion
A THz metamaterial based on toroidal dipole resonance is explored by simulations and experiment.It is shown that aQfactor of 26.6 and a refractive index sensitivity of 337.5 GHz/RIU can be achieved.For the AFB1 biosensors in our experiment, dip amplitudes of transmission varying from 0.1904 to 0.203 and 0.2093 are observed as AFB1 concentrations are altered from 0 to 0.001 µg·ml-1and 0.01 µg·ml-1respectively.When AFB1 concentrations are 0.1 µg·ml-1, 1 µg·ml-1, 10 µg·ml-1, 100 µg·ml-1, dip amplitudes of 0.2179,0.226,0.2384 and 0.2527 and dip redshifts of 10.1 GHz, 20.1 GHz, 27.7 GHz and 37.6 GHz are respectively obtained.AFB1 shows strong absorption of THz waves.In the experiment,the change in transmission spectrum amplitude of metamaterials is mainly caused by the absorption of AFB1,and the redshift of resonant frequency is influenced by the different thicknesses of the analyte layer on the device surface.The experimental results imply that the minimum measurable concentration can be as low as 0.001 µg·ml-1, illustrating high-sensitivity, label-free detection of AFB1 and enriching the applications of sensors in the THz domain.
Appendix A:Finite-element simulations
The designed structure is simulated with CST Microwave Studio.All parameters of the metamaterial structures follow the designed structures, as shown in Fig.1.We note that in the simulations the electric conductivity of copper is set asσ=5.8×107S·m-1.Periodic boundaries were used in thexandydirections;thezdirection was set to an open boundary.
Appendix B:AFB1 preparation and detection
The 100 µg·ml-1AFB1 standard solution sample was purchased from Qingdao Lihai Pharmaceutical Packaging Technology Co., Ltd.The AFB1 standard solution was prepared with ethanol at concentrations of 100 µg·ml-1,10 µg·ml-1, 1 µg·ml-1, 0.1 µg·ml-1, 0.01 µg·ml-1and 0.001µg·ml-1.Ten microliters of each solution were dropped on the surface of the toroidal dipole sensors.Subsequently,the samples were dried at room temperature for 3 min.These two steps were repeated three times.The transmission spectra of all samples were measured by a THz-TDS filled with nitrogen.
Acknowledgements
Project supported by the National Natural Science Foundation of China (Grant Nos.61927813, 61865009, and 12104203)and Jiangxi Provincial Natural Science Foundation(Grant No.20212ACB201007).
猜你喜欢
杂志排行
Chinese Physics B的其它文章
- Does the Hartman effect exist in triangular barriers
- Quantum geometric tensor and the topological characterization of the extended Su–Schrieffer–Heeger model
- A lightweight symmetric image encryption cryptosystem in wavelet domain based on an improved sine map
- Effects of drive imbalance on the particle emission from a Bose–Einstein condensate in a one-dimensional lattice
- A new quantum key distribution resource allocation and routing optimization scheme
- Coexistence behavior of asymmetric attractors in hyperbolic-type memristive Hopfield neural network and its application in image encryption