Bedside ultrasonic localization of the nasogastric tube in a patient with severe COVID-19: A case report

2022-06-22XiaoJuZhuShuiXiaLiuQiuTangLiYuanJingJiang

World Journal of Clinical Cases 2022年15期

INTRODUCTION

Coronavirus disease 2019 (COVID-19) is a novel acute respiratory infectious disease caused by severe acute respiratory syndrome coronavirus 2. COVID-19 is characterized by a rapid rate of transmission and susceptibility of the population. The disease usually manifests with symptoms of fever, dry cough,and fatigue. In severe cases, patients may experience dyspnea and/or hypoxemia in the week following onset. Severe COVID-19 can rapidly progress to acute respiratory distress syndrome, septic shock,irreversible metabolic acidosis, coagulation dysfunction, and multiple organ failure[1-2]. Due to reduced food intake, increased energy consumption, and impaired anabolic functioning, the incidence of malnutrition in these patients can reach 25% to 65%[3], and may adversely affect patient outcomes.Early enteral nutrition is considered one of the active treatment measures for critically ill patients as it can reduce the severity of the disease and the incidence of complications, as well as shorten hospitalization and improve patient prognosis[4-6].

All the country people were standing10 round the King s throne, and were crowded together in thick masses almost out of the windows to see the Princess receive the suitors; and as each one came into the room all his fine phrases went out like a candle! It doesn t matter! said the Princess

The indwelling nasogastric tube is the most commonly used method for supplying enteral nutrition.The tube requires accurate positioning. Traditional methods include extraction and examination of the patient’s gastric juice, followed by determination of the sound of air passing through the liquid. The tube can also be positioned

using ultrasound and X-rays. Under COVID-19 and Level 3 protection requirements, the use of bedside ultrasound is advantageous in situations where the stethoscope cannot be used, and it does not require extensive human and material resources.

CASE PRESENTATION

Chief complaints

A 79-year-old patient was admitted to the Department of Critical Care Medicine, the COVID-19 Specialist Hospital of Wuhan Taikang Tongji Hospital (Wuhan, China), with severe COVID-19, coronary atherosclerotic heart disease, and cerebral infarction in March 2020.

History of present illness

The patient had developed fever without obvious inducement 7 wk prior, with a body temperature reaching 38.5 °C, fatigue and anorexia. The body temperature then fluctuated between 36.5 °C and 38°C. Since the onset of the disease, the patient had had no contact with COVID-19 patients. At the time of admission, the patient was severely ill, complicated by pulmonary bacterial infection, respiratory failure, co-morbid diseases, and poor nutrition. The patient required a nasogastric tube for enteral nutritional support, which was inserted in the hospital under medical supervision.

History of past illness

The patient had a history of cerebral infarction and recovered through drug treatment without obvious impairment of either action or speech.

Personal and family history

There was no history of COVID-19 or other medical issues in the patient‘s family.

Physical examination

On arrival in the department, the nutritional status of the patient and the feasibility of nasal feeding of the patient were assessed. In addition, the patient was investigated for nasal septum deviation, nasal inflammation, obstruction, cerebrospinal fluid rhinorrhea, and other conditions that would render indwelling unsuitable.

Laboratory examinations

After admission, laboratory examinations were conducted and the patient received anti-infection and antiviral treatment, together with treatments for improving cardiac function, blood transfusions, and other symptomatic treatments.

Imaging examinations

The patient was severely ill, complicated by pulmonary bacterial infection, respiratory failure, comorbid diseases, and poor nutrition. The patient required a nasogastric tube for enteral nutritional support.

FINAL DIAGNOSIS

An initial imaging evaluation with brain computed tomography (CT) scan suggested multiple small punctate hypodense foci next to the lateral ventricles bilaterally, with ischemic changes and a high probability of acute ischemic stroke; the chest CT revealed scattered patches and cloudy fuzzy shadows in both lungs, and the patient was considered to have been infected with COVID-19.

TREATMENT

The required length of the nasogastric tube was determined and marked during physical examination.Then the tube was inserted to the predetermined length by the routine procedure. The nasogastric tube was gently placed to avoid damaging the esophageal mucosa and to ensure that the tube was not coiled in the mouth.

OUTCOME AND FOLLOW-UP

The gastric juice was not able to be extracted using an empty needle. Meanwhile, due to the particular circumstances, the stethoscope could not be used for auscultation. The end of the nasogastric tube was placed in normal saline solution without bubble overflow. As a result, it was unknown whether the tube was in the stomach, and no nutrient solution was injected. The tube position was then examined

ultrasound. The ultrasonic probe could clearly identify the circular nasogastric tube and comet tail sign through the cross-section after the esophageal inlet. On examination of the longitudinal section of the esophagus, it was observed that the nasogastric tube was running parallel in the esophagus, seen as two parallel transparent hyperechoic images in the shape of “=” (Figure 1). This indicated the position of the nasogastric tube, and then 100 mL nutrient solution was injected. The patient suffered no discomfort and the vital signs were stable without adverse reactions.

The end of the tube is placed in a bowl filled with normal saline to determine overflow of the bubbles. However, bubble overflow may only indicate that the tube is in the airway and is not able to detect whether it is in the stomach. It is also not recommended for routine use as it may increase the risk of inadvertent aspiration when the patient is aspirating[8].

DISCUSSION

Routine diagnostic methods

: Conventionally, a stethoscope with a 50 mL disposable syringe is used for auscultation. Initially, 20 to 30 mL of air is rapidly injected, followed by repeated auscultation in the abdomen. Detection of a high-profile sound of air passing through the liquid indicates that the tube is located in the stomach.

If you want something badly enough,You must let it go free.If it comes back to you,It s yours.If it doesn t,You really never had it anyway.——Anonymous

A sterile syringe is connected to the end of the tube to determine whether gastric juice can be extracted.

The end of the tube is placed in a bowl filled with cold boiled water or normal saline to determine the presence of bubbles.

A bedside X-ray film is taken within 2 h after insertion of the nasogastric tube, and mobile digital radiography is performed. The radiologist obtains anteriorposterior chest images according to the specifications and completes the film reading and the report.

At last the young man gave in, and threw the fish back into the sea; and both brothers went supperless to bed, and wondered what fortune the next day would bring

The gas status in the patient's stomach should be identified. If a strong echo appears with too much gas, this may affect the judgment of the result. Thus, it is necessary to extract the gas before exploration. After inserting the nasogastric tube, the circular tube and the comet tail sign were clearly seen in the transverse section after the ultrasonic probe passed through the esophageal inlet. In the longitudinal section, the nasogastric tube was found to run parallel in the esophagus, seen as two parallel transparent hyperechoic images in the shape of “=”.

At last the woman came back again, and said in a gruff voice, Good evening, Zachiel; when the young moon shines in the basket, you are freed early, Zachiel

Disadvantages of conventional diagnostic methods and advantages of ultrasound diagnosis

The authors would like to thank the patient and his family for their cooperation in relation to this case report.

Clinically, we found that most critical patients with COVID-19 are unable to eat. After insertion of the nasogastric tube, adhesion and reverse folding of the tube may occur, resulting in too little liquid in the stomach. As a result, it is impossible to locate and extract the nasogastric contents accurately, thus limiting the efficacy of the gastric juice extraction method.

Thirsty as he was, he did not wait to drink, but he told his mother that he smelt the blood of a Christian man, and that she had better bring him out at once and make him ready to be eaten

The determination of nasogastric tube positioning by bedside radiography is the gold standard for determining accurate positioning; however, this has the disadvantages of high levels of radiation and the length of the procedure, making it unsuitable for repeated use[9-10]. In addition, with COVID-19, all medical personnel are required to use Grade 3 protection, and the removal of instruments from the radiology department not only increases the workload for the radiology personnel but also affects the timing and efficiency of the treatment. Also,when performing abdominal plain X-rays for severely ill patients, usually at least three medical personnel are required to move and place the body in position, resulting in increased physical exertion.Furthermore, when moving, the radius of action would tend to increase, as would the amount of perspiration from the medical personnel, which would tend to reduce the effectiveness of the protective clothing and thus increase the risk of exposure. For patients with severe disease, there are usually many tubes attached. The risk of tube slippage would increase and thus the quality of nursing would tend to be reduced during movement.

Chongqing Medical Scientific Research Project, No. 2020FYYX046

CONCLUSION

Accurate positioning of the nasogastric tube in severely ill patients with COVID-19 using bedside ultrasound could reduce both the risk of viral exposure to medical personnel and the clinical workload,as well as avoid X-ray radiation. Bedside ultrasound localization of the nasogastric tube produces immediate accurate results, thus gaining valuable time for clinical treatment and effectively improving both the potential success of the treatment and prognosis of the patient. It is thus worthy of clinical promotion and application.

They looked to him exactly like the ones he had left, except that they moved more slowly still, and each one held a raven63 upon his wrist, and their harsh croakings re-echoed through the dismal64 place

ACKNOWLEDGEMENTS

When air is injected into the nasogastric tube, the sound of air passing through liquid is strongly discernable on auscultation. The sound of air injected into the trachea and pleural cavity is not easy to distinguish from that injected into the stomach. Therefore, it is not easy to accurately distinguish the location of the nasogastric tube by auscultation, and some patients may present with clinical manifestations such as choking and decreased oxygen saturation[7]. Given that COVID-19 is a Class B infectious disease and can be transmitted by contact and droplets, medical personnel need Level 3 protection when entering the Red zone to reduce the risk of infection. Additionally, a stethoscope used in the Red zone becomes a contaminated object, and viruses and other pollutants may attach to the surface. Because of this, the stethoscope should not come into direct contact with the body surface of the medical personnel. In addition, the senses, including sensation in the limbs, and especially hearing, are greatly weakened, which adversely affect the accuracy of judgment and auscultation. Therefore,conventional auscultation in patients with COVID-19 increases the risk of exposure of medical personnel to the virus and is also difficult to implement.

FOOTNOTES

Zhu XJ was the patient’s supervisor nurse, who carried out the study, reviewed the literature,contributed to manuscript drafting, and was responsible for revising the manuscript for important intellectual content; Liu SX, Li QT, and Jiang YJ reviewed the literature and contributed to manuscript drafting; All authors issued final approval for the version to be submitted.

Studies have reported a 100% accuracy rate of bedside ultrasound in determining the success of gastric tube placement[11].Ultrasound is a non-invasive procedure with the advantages of repeatability and a lack of radiation[12]. For unconscious patients, bedside ultrasound positioning is safe and can be carried out with other diagnostic and treatment measures without affecting mechanical ventilation, delaying treatment, and transferring risk. Furthermore, bedside ultrasound has been reported to intuitively indicate the patient's condition and pathology, improve the precision of clinical diagnosis, guide the positioning of intubation and puncture operation and therefore reduces damage to the laryngeal mucosa caused by repeated manipulation[13]. Moreover, given its intuitive, fast, accurate, and convenient features, bedside ultrasound has been applied as a standard tool for monitoring and evaluating severely ill patients and positioning of punctures in intensive care unit,anesthesia, and emergency departments in numerous countries[14-16]. It is not difficult to locate the position of the nasogastric tube if the thyroid and tracheal positions are located. In addition, there is less disturbance to the neck. If the position of the nasogastric tube cannot be determined due to pseudointerference or other causes, about 20 mL of air should be quickly injected into the tube. This would result in the formation of a prominent gas reflection plane, and the position of the tube would be visible as a strong strip echo under ultrasound. The intensive care unit of our department has an ultrasound machine to reduce the increased workload caused by repeated demand. This instrument is always available to facilitate the timely diagnosis and treatment of patients.

No sooner was this done than the Wizard King was seen in the air under the form of some unknown bird, exclaiming as he flew off that he would never forgive either his son or the Fairy the cruel wrong they had done him

All study participants, or their legal guardian, and the patient has provided informed written consent for publication of this case report and accompanying images prior to study enrollment.

The authors have no conflicts of interest to declare.

The authors have read the CARE Checklist (2016), and the manuscript was prepared and revised according to the CARE Checklist (2016).

This article is an open-access article that was selected by an in-house editor and fully peer-reviewed by external reviewers. It is distributed in accordance with the Creative Commons Attribution NonCommercial (CC BYNC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited and the use is noncommercial. See: https://creativecommons.org/Licenses/by-nc/4.0/

China

The Cat s master did not build very much upon what he said. He had often seen him play a great many cunning tricks to catch rats and mice,15 as when he used to hang by the heels, or hide himself in the meal, and make as if he were dead; so that he did not altogether despair of his affording him some help in his miserable condition. When the Cat had what he asked for he booted himself very gallantly, and putting his bag about his neck, he held the strings of it in his two forepaws and went into a warren16 where was great abundance of rabbits. He put bran and sow-thistle into his bag, and stretching out at length, as if he had been dead, he waited for some young rabbits, not yet acquainted with the deceits of the world, to come and rummage his bag for what he had put into it.

Xiao-Ju Zhu 0000-0001-7257-3818; Shui-Xia Liu 0000-0003-3472-6216; Qiu-Tang Li 0000-0002-3128-1293;Yuan-Jing Jiang 000-0002-0204-6235.

Xing YX

Filipodia

After that, my friends kept pushing me: Why don t you ask him for one of his home-run bats? Finally, when Roger was standing7 by the fence, I made the request. He said, Sure. Next time I break one.

Xing YX

1 Wu Z, McGoogan JM. Characteristics of and Important Lessons From the Coronavirus Disease 2019 (COVID-19)Outbreak in China: Summary of a Report of 72 314 Cases From the Chinese Center for Disease Control and Prevention.

2020; 323: 1239-1242 [PMID: 32091533 DOI: 10.1001/jama.2020.2648]

2 Wiersinga WJ, Rhodes A, Cheng AC, Peacock SJ, Prescott HC. Pathophysiology, Transmission, Diagnosis, and Treatment of Coronavirus Disease 2019 (COVID-19): A Review.

2020; 324: 782-793 [PMID: 32648899 DOI:10.1001/jama.2020.12839]

3 Zhu M, Wu WR. Enteral nutrition. Beijing: People’s Health Publishing House, 2002: 377-380.

4 Guo YB, Liu Y, Ma J, Cai Y, Jiang XM, Zhang H. Effect of early enteral nutrition support for the management of acute severe pancreatitis: A protocol of systematic review.

2020; 99: e21569 [PMID: 32769901 DOI:10.1097/MD.0000000000021569]

5 Ge W, Wei W, Shuang P, Yan-Xia Z, Ling L. Nasointestinal Tube in Mechanical Ventilation Patients is More Advantageous.

2019; 14: 426-430 [PMID: 31198856 DOI: 10.1515/med-2019-0045]

6 Jiang Z, Wen C, Wang C, Zhao Z, Bo L, Wan X, Deng X. Plasma metabolomics of early parenteral nutrition followed with enteral nutrition in pancreatic surgery patients.

2019; 9: 18846 [PMID: 31827206 DOI:10.1038/s41598-019-55440-z]

7 Tsung JW, Fenster D, Kessler DO, Novik J. Dynamic anatomic relationship of the esophagus and trachea on sonography:implications for endotracheal tube confirmation in children.

2012; 31: 1365-1370 [PMID: 22922616 DOI: 10.7863/jum.2012.31.9.1365]

8 Roberts S, Echeverria P, Gabriel SA. Devices and techniques for bedside enteral feeding tube placement.

2007; 22: 412-420 [PMID: 17644695 DOI: 10.1177/0115426507022004412]

9 Motz P, Von Saint Andre Von Arnim A, Iyer RS, Chabra S, Likes M, Dighe M. Point-of-care ultrasound for peripherally inserted central catheter monitoring: a pilot study.

2019; 47: 991-996 [PMID: 31605580 DOI:10.1515/jpm-2019-0198]

10 Telang N, Sharma D, Pratap OT, Kandraju H, Murki S. Use of real-time ultrasound for locating tip position in neonates undergoing peripherally inserted central catheter insertion: A pilot study.

2017; 145: 373-376 [PMID:28749401 DOI: 10.4103/ijmr.IJMR_1542_14]

11 Zatelli M, Vezzali N. 4-Point ultrasonography to confirm the correct position of the nasogastric tube in 114 critically ill patients.

2017; 20: 53-58 [PMID: 28298944 DOI: 10.1007/s40477-016-0219-0]

12 Heidegger CP, Graf S, Perneger T, Genton L, Oshima T, Pichard C. The burden of diarrhea in the intensive care unit (ICUBD). A survey and observational study of the caregivers' opinions and workload.

2016; 59: 163-168 [PMID:27222461 DOI: 10.1016/j.ijnurstu.2016.04.005]

13 Hirose T, Shimizu K, Ogura H, Tasaki O, Hamasaki T, Yamano S, Ohnishi M, Kuwagata Y, Shimazu T. Altered balance of the aminogram in patients with sepsis - the relation to mortality.

2014; 33: 179-182 [PMID: 24377412 DOI:10.1016/j.clnu.2013.11.017]

14 Lichtenstein D. Fluid administration limited by lung sonography: the place of lung ultrasound in assessment of acute circulatory failure (the FALLS-protocol).

2012; 6: 155-162 [PMID: 22455488 DOI:10.1586/ers.12.13]

15 Wang G, Ji X, Xu Y, Xiang X. Lung ultrasound: a promising tool to monitor ventilator-associated pneumonia in critically ill patients.

2016; 20: 320 [PMID: 27784331 DOI: 10.1186/s13054-016-1487-y]

16 Xirouchaki N, Kondili E, Prinianakis G, Malliotakis P, Georgopoulos D. Impact of lung ultrasound on clinical decision making in critically ill patients.

2014; 40: 57-65 [PMID: 24158410 DOI: 10.1007/s00134-013-3133-3]