APP下载

Evaluation of internal and shell stiffness in the differential diagnosis of breast non-mass lesions by shear wave elastography

2020-09-15

World Journal of Clinical Cases 2020年12期

Ping Xu,Min Yang,Lan-Ying Wu,Department of Ultrasound,Beijing Shijitan Hospital Affiliated to Capital Medical University,Beijing 100038,China

Mei Wu,Zheng-Min Ruan,Department of Ultrasound,The Second Hospital of Shandong University,Jinan 250033,Shandong Province,China

Juan Xiao,Evidence-based Medicine Center,The Second Hospital of Shandong University,Jinan 250033,Shandong Province,China

Abstract

Key words:Breast tumor;Shear wave elastography;Non-mass lesions;Stiff rim sign;Ultrasound;Diagnosis

INTRODUCTION

Breast cancer is the most common malignant tumor in women and one of the most common causes of death[1].Timely and accurate diagnosis is crucial for treatment and prognosis[2].Breast lesions can be classified as mass lesions and non-mass lesions(NMLs)[3].NMLs are characterized by no clear boundary and no space occupying effect in two or more different scanning sections on gray-scale ultrasonography.The diagnostic specificity of conventional ultrasound is low at approximately 21%-43%[4-6].Shear wave elastography (SWE) can distinguish benign lesions from malignant lesions by evaluating the stiffness of breast masses,SWE has good reproducibility[7]and high diagnostic efficacy[8-10].However,there are very few independent studies on the diagnostic value of SWE in breast NMLs[5,11].Studies have shown that the hardest areas of malignant lesions were located in the periphery rather than the interior of breast lesions,which means that elasticity analysis of the periphery may have good diagnostic value[12].Qualitative and quantitative analysis of elasticity in the surrounding tissues of NMLs was performed in our study innovatively[13].In the present study,118 patients with NMLs underwent SWE examination to analyze the elasticity characteristics and their clinical diagnostic values.

MATERIALS AND METHODS

Patients

From January 2019 to January 2020,118 patients with breast NMLs were enrolled in this study in Beijing Shijitan Hospital and The Second Hospital of Shandong University.All patients were female,aged 24 to 68 years.All patients who underwent breast NMLs biopsy or surgery signed an informed consent.

Inclusion criteria:(1) Conventional ultrasonography shows the characteristics of NMLs[3];(2) Diagnosis was confirmed by pathological results;and (3) Conventional ultrasonography and SWE examination were performed before surgery.

Exclusion criteria:(1) Having accepted radiotherapy or endocrine therapy before examination;(2) Surgical scar within the lesion area;(3) Cystic or cystic-solid mixed mass[14];(4) Lesion maximum diameter>4.0 cm;and (5) Massive calcification within the lesion area.

Conventional ultrasound and SWE examination

Conventional ultrasound and SWE examination were performed using Ultrasonic System Resona 7 (Mindray,Shenzhen,China),with a line array transducer (L14-5W),frequency 4.0-14.0 MHz.Examination method:Conventional ultrasound examination of the breast was performed first,and then switched to SWE mode.The section with the largest lesion was selected,and the lesion was placed in the center of the sampling box.SWE mass-velocity dual dynamic pattern was used to perform the examination.Patients were asked to hold their breath.A good quality “stiff rim” sign was observed,which showed that the mass diagram on the left of the screen had a homogeneous green background without an obvious pseudo-image in purple[15](Figure1).In the right image on the screen,red indicated “hard tissue” and blue indicated “soft tissue”.The lesion with a red or orange ring was seen as the “stiff rim”sign,indicating that the lesion was malignant.The lesion without a “stiff rim” sign was thought to be benign.Tracing the outer edge of the lesion on gray-scale image,the internal elastic parameters,including maximum (Emax),mean (Emean),minimum and standard deviation (Esd) of NMLs were calculated automatically in gray-scalevelocity dual pattern.Then,Emax and Emean elasticity values at shells of 1 mm,1.5 mm,2 mm,2.5 mm and 3 mm from the edge of NMLs were measured,respectively(Figure2 and 3).All images and data were recorded on the internal hard disk of the ultrasonic system.All examinations were performed by two radiologists both with more than 10 years of experience on breast diagnoses.They analyzed the ultrasonography independently.If their opinions differed,an agreement was reached by discussion.

Statistical analysis

Pathological results were the gold standard.SPSS22.0 software was used for statistical analysis.Quantitative parameters of SWE are shown as mean ± standard deviation and comparisons between groups were evaluated using thettest.Qualitative parameters are shown as frequency or percentage and comparisons between groups were evaluated using theχ2test.A receiver operating characteristic (ROC) curve was plotted and then the areas under the curve,Youden indices (sensitivity + specificity -1),cutoff values and the corresponding sensitivity,specificity and accuracy of each parameter were calculated.P< 0.05 was considered statistically significant.

RESULTS

General information

There were no significant differences in age,maximum diameter,location,amenorrhea and history of lactation between the two groups (Table1).

Pathological results

Of the 118 patients,66 were diagnosed with benign lesions and 52 with malignant lesions.There were 30 cases of intraductal carcinoma in situ,19 cases of invasive ductal carcinoma,2 cases of invasive lobular carcinoma and 1 case of lymphoma in the malignant group.There were 32 cases of adenosis,9 cases of intraductal papilloma,8 cases of breast inflammation,7 cases of atypical ductal hyperplasia,6 cases of fibroadenoma and 4 cases of sclerosing adenosis in the benign group.

Comparison of elasticity parameters in benign and malignant NMLs

Comparison of internal elasticity parameters in benign and malignant NMLs:Emax,Emean and Esd in malignant NMLs were significantly higher than those in benign lesions (P< 0.05),and there was no significant difference in minimum value between the two groups (Table2).

Comparison of elasticity parameters around NMLs:The percentage of “stiff rim”signs in malignant NMLs was significantly higher than that in benign lesions (P<0.05).Emax and Emean of the shell at 1 mm,1.5 mm,2 mm,2.5 mm and 3 mm in the malignant group were also higher than those in the benign group (P< 0.05) (Table3).

ROC curves of each parameter and corresponding sensitivity,specificity,accuracy,cutoff value and the area under the curve of each parameter:The ROC curve of each parameter are shown in Figure4.The sensitivity,specificity,accuracy,cutoff value and the areas under the curve (AUC) are presented in Table4.Emax at the shell of 2.5 mm in NMLs had the maximum AUC of 0.900,and the corresponding sensitivity,specificity accuracy and cutoff value were 94.57%,85.86%,87.44% and 94.62 kPa,respectively.

Figure1 Mass-velocity dynamic dual pattern.Homogeneous green background is shown on the left image.The“stiff rim” sign (a ring of red and orange round the lesions) is shown on the right image.RLB index:Reliable index.

DISCUSSION

Previous research reported that both benign and malignant lesions could present as NMLs[3,16,17],accounting for approximately 9.2% of all breast lesions[18].The diagnostic specificity of conventional ultrasound is low in distinguishing malignant lesions from benign lesions[4].Some studies[5,19]have shown that SWE had better diagnostic performance than conventional ultrasound in the diagnosis of benign and malignant breast NMLs;however,these studies mainly focused on the internal elasticity characteristics of the lesions.In addition,some studies[13]suggested that the stiffness evaluation of surrounding tissues of breast lesions might have good differential diagnostic value.The elasticity characteristics and diagnostic value of the stiffness of the surrounding tissues in NMLs are still not clear,and there is no independent studies on the qualitative and quantitative evaluation of these parameters.

In the present study,we found that the internal and peripheral Emax and Emean of malignant NMLs were higher than those of benign NMLs,which is consistent with a previous study on breast mass lesions[20].However,the mean values of internal elasticity parameters of NMLs were lower than those of mass lesions in the malignant group,and so was the diagnostic efficiency of internal parameters.This was mainly related to the high percentage of intraductal carcinoma in situ in malignant NMLs.Our results indicate that evaluation of the surrounding tissues may be more helpful in the differential diagnosis of NMLs than internal SWE elasticity.The hardest tissue with pathological changes was not located inside the lesion but on the peripheral tissue,which was mainly because malignant lesions tend to seep into the surrounding tissue[12].Cancer cells infiltrate and diffuse to the peripheral interstitial tissue,and this dynamic reaction leads to the formation of the mixed zone.The mixed zone may consist of cell proliferation,lymphocytic infiltration,fibrosis,and angiogenesis of the tumor[13],which could increase the peripheral stiffness around malignant lesions up to 2-10 times higher[21].Another study[22]reported that “stiff rim” sign was caused by attenuation of the energy of the shear wave of the peritumoral region of the lesion.

Figure2 Gray-scale-velocity dual pattern.The traced line of the outer edge (white circle) of the NMLs and the shell at 2.5 mm (pink circle) can be seen on the left image.A “stiff rim” of red and orange can also be seen on the right image.Mean,maximum,and minimum elastic values and the standard deviation were 33.80 kPa,92.869 kPa,6.23 kPa and 14.41 kPa,respectively,for internal elasticity,and 81.15 kPa,167.97 kPa,5.03 kPa and 28.05 kPa,respectively,for the shell at 2.5 mm.The pathology result was infiltrating ductal carcinoma.A:The area within the lesion.Shell:The area around the lesion.Emax:Maximum;Emean:Mean;Min:Minimum;Esd:Elastic values and the standard deviation.

Qualitative and quantitative analyses of the stiffness parameters in the tissues surrounding NMLs,including the qualitative parameter (“stiff rim” sign) and quantitative parameters (elasticity values of the shell at 1 mm,1.5 mm,2 mm,2.5 mm and 3 mm around the lesion) were performed.Using shell function,we found that the presence of the “stiff rim” sign in the malignant group was significantly higher than that in the benign group.In peripheral stiffness quantitative analysis,Emax and Emean of the shell within 1-3 mm were significantly higher in the malignant group than in the benign group.These results showed that elasticity analysis of the surrounding tissues of NMLs may have the potential ability to distinguish benign lesions from malignant lesions.The “stiff rim” sign and Emax/Emean of the shell within 1-3 mm have high diagnostic value,and the AUC of Emax at the shell of 2.5 mm has the largest AUC of 0.900.Huanget al[13]studied the surrounding tissues of breast mass lesions and found that Emax of the shell at 3 mm had the best diagnostic value.Possible reasons for this are as follows:(1) In our study,there was a large percentage of intraductal carcinoma in situ,which has mild invasion ability to the surrounding region;and (2) Our study thins out the shell layer.Of all the quantitative elasticity parameters,Emax had better diagnostic efficacy than the others,which was similar to previous studies[15,23].This was mainly because Emax,Emean and Esd reflected different pathological anatomy changes.Emean reflects the average stiffness in the region of interest (ROI),and Esd represents the heterogeneity of various tissue textures in the ROI.Cancers are histologically heterogeneous due to the heterogeneity of the tumor cell population and the cancer microenvironment.Emean and Esd were more likely affected by internal liquefaction,necrosis,calcification,collagenetc.Emax reflects the maximum stiffness of the hardest tissue in the ROI,which is rarely affected by other factors.Huanget al[24]reported that Esd can improve the diagnosis sensitivity of malignant lesions.The different results obtained in different studies might be due to the different instruments used and the ROI type.Although Emean,Esd and Emax are all stiffness parameters,the selection of appropriate parameters is important,as they reflect different pathological formations.

SWE had good diagnostic value in NMLs;however,the false negative and false positive results should not be ignored.Many studies[6,25,26]have shown that multiple clinical and ultrasonic factors are associated with false negative and false positive results.In malignant NMLs,calcification,small lesion size and the appearance of in situ carcinoma were associated with false negative results.In benign NMLs,short distance to the nipple was the main reason for false positive results[6].To reduce both false positive and false negative findings,it is necessary to make a diagnosis using a combination of other imaging techniques,for instance,mammography is necessary in cases with calcification in NMLs.

As the internal and peripheral parameters reflect different pathological formations of the lesion,combined application of imaging techniques could further improve diagnostic accuracy in NMLs.The combination of qualitative and quantitative analyses of stiffness in the circumjacent area around NMLs was applied in this study,which showed that Emax yielded a better performance than Emean as well as the elasticity parameters of surrounding tissues.Our study still has some limitations.The sample size was small.Qualitative and quantitative analyses of stiffness in the circumjacent area of NMLs by SWE still requires further investigation in a multicenter trial with a large sample size.

In conclusion,the “stiff rim” sign and other quantitative parameters within and around the breast NMLs have good diagnosis accuracy.The Emax of peripheral shells had the best evaluation efficiency.Combining qualitative and quantitative analyses of stiffness may further improve the accuracy of diagnosis.

Table1 Comparison of general information between the benign group and the malignant non-mass lesions group

Table2 Comparison of internal elasticity parameters of non-mass lesions (kPa)

Table3 Comparison of elasticity parameters around non-mass lesions

Table4 Sensitivity,specificity,accuracy,cutoff value and the area under the curve of each parameter

Figure3 Gray-scale-velocity dual pattern.This shows the traced line of the outer edge (white circle) of the NMLs and the shell at 2.5 mm (pink circle) on the left image,and no “stiff rim” on the right image.Mean,maximum,minimum elastic values and the standard deviation were 17.03 kPa,40.89 kPa,3.23 kPa and 6.93 kPa,respectively,for internal elasticity,and 12.74 kPa,37.00 kPa,2.51 kPa and 7.90 kPa,respectively,for the shell at 2.5 mm.The pathology result was adenosis.A:The area within the lesion;Shell:The area around the lesion.Emax:Maximum;Emean:Mean;Min:Minimum;Esd:Elastic values and the standard deviation.

Figure4 Receiver operating characteristic curves for internal and peripheral elasticity parameters of the shell at 1 mm,1.5 mm,2 mm,2.5 mm,and 3 mm,respectively.A:Receiver operating characteristic curves for internal elasticity parameters maximum (Emax),mean (Emean),minimum (Emin) elastic values and the standard deviation (Esd),the area under the curve (AUC) was 0.707,0.626,0.425 and 0.687,respectively;B:Emax and Emean of the shell at 1 mm (as E1max and E1mean),AUC was 0.795 and 0.647 respectively;C:Emax and Emean of the shell at 1.5 mm (as E1.5max and E1.5mean),AUC was 0.789 and 0.692,respectively;D:Emax and Emean of the shell at 2 mm (as E2max and E2mean),AUC was 0.826 and 0.704,respectively.E:Emax and Emean of the shell at 2.5 mm (as E2.5max and E2.5mean),AUC was 0.900 and 0.746,respectively;F:Emax and Emean of the shell at 3 mm (as E3max and E3mean),AUC was 0.881 and 0.734,respectively.Emax:Maximum;Emean:Mean;Min:Minimum;Esd:Elastic values and the standard deviation;AUC:The area under the curve.

ARTICLE HIGHLIGHTS

Research background

Shear wave elastography can help the differential diagnosis of breast lesions by evaluating the hardness of breast lesions;however,there are few studies on the evaluation of the elasticity of breast non-mass lesions.In recent years,studies have shown that the evaluation of peripheral elasticity of breast lesions may have better diagnostic efficacy than other parameters.However,there has been no study on the internal and peripheral elasticity of breast non-mass lesions.

Research motivation

The aim of this study was to evaluate the internal and peripheral elasticity of breast non-mass lesions using shear wave elastography.

Research objectives

To determine the value of shear wave elastography in the diff erential diagnosis of non-mass breast lesions.

Research methods

The peripheral (the shell of 1 mm,1.5 mm,2 mm,2.5 mm and 3 mm around the lesions) and internal elasticity of non-mass breast lesions in 118 cases were evaluated.ROC curves of each parameter were drawn and the diagnostic efficacy of each parameter was compared.

Research results

The “stiff rim” sign and other quantitative parameters within and around the breast NMLs had good diagnostic efficiency.Emax of peripheral shells had better evaluation efficiency.

Research conclusions

Combining qualitative and quantitative analyses of both internal and shell stiffness may further improve the diagnostic efficiency of breast non-mass lesions.

Research perspectives

In this study,we focused on the influence of the stiffness of the peripheral shell in breast nonmass lesions on diagnosis.