心脏除颤器测量不确定度评定
2024-10-12邝仕敏
【摘要】心脏除颤器作为一种急救类医疗设备,广泛应用于各医疗场所。其各项计量性能指标的准确性不仅影响急救的效果,还与病人的身体健康安全息息相关。本文针对该类设备相应的参数进行测量不确定度的分析与评定,以提高校准结果的准确性,为相关计量检定人员提供参考。
【关键词】心脏除颤器;不确定度;示值误差
【DOI编码】10.3969/j.issn.1674-4977.2024.05.077
Evaluation of Measurement Uncertainty for Defibrillators
KUANG Shimin
(Hainan Provincial Inspection Institute,Haikou 570206 China)
Abstract: As emergency medical devices, defibrillators are widely used in various medical settings. The accuracy of their various metrological performance indicators not only affects the effectiveness of emergency treatment but is also closely related to the health and safety of patients. This article analyzes and evaluates the measurement uncertainty of the corresponding parameters for such devices to enhance the accuracy of calibration results, providing a reference for metrological verification personnel.
Keywords: defibrillator; uncertainty; indication error
1.释放能量不确定度的评定
根据校准规范,除颤分析仪释放能量最大允许误差为±2 J或测量值的±5%,两者取大值。因此,在心脏除颤器0~400 J的测量范围内应分两部分来进行释放能量的不确定度分析与评定。
1.1在(0,40] J测量范围内释放能量的测量不确定度评定

1.1.1测量不确定度的分量
1)由心脏除颤器释放能量测得值重复性引入的不确定度分量
在除颤分析仪(0,40] J的测量范围内,将被校仪器释放能量挡选择为40 J,在该能量点测量10次释放能量,测量结果分别为39.8、39.6、39.7、39.2、39.5、39.9、39.6、39.3、39.7、39.5 J。计算得单次测量结果的实验标准偏差为s( E ) = 0.22J。因校准过程中所需测量值为单次测量的结果,故所引入标准不确定度分量为:





【参考文献】
[1]心脏除颤器校准规范:JJF 1149—2014[S].
[2]测量不确定度评定与表示:JJF 1059.1—2012[S].
[3]古丽孜拉.心脏除颤器和心脏除颤监护仪测量不确定度评定[J].计量与测试技术,2015,42(8):78-80.
【作者简介】
邝仕敏,男,1988年出生,工程师,学士,研究方向为医学计量检定。
(编辑:李加鹏)